User Manual
Calibration Procedure
To calibrate the detector perform the following procedure.
Note
Calibrate O
2
in clean air.
If C is pressed at any time during the calibration
procedure, calibration is ABORTED. The detector
emits a quick beep and the calibration due dates
are displayed before the detector returns to normal
operation.
Start Calibration
Note
Verify that the calibration gas you are using
matches the span concentration value(s) in the
detector. See the section,
Calibration Gas
Concentration.
1. In a clean atmosphere, press and hold
C as the
detector beeps, flashes, and vibrates to the
corresponding OFF countdown. Continue to hold
C as the detector briefly deactivates.
2. The detector then reactivates and performs the
CAL countdown. Continue to hold C until the CAL
countdown is complete to enter calibration.
Note
If C is not held for the entire CAL countdown, the
detector will deactivate.